This store requires javascript to be enabled for some features to work correctly.

Sale

Drager Evita XL Ventilator

  • $ 5,450.00

Drager Evita XL Ventilator

Neonatal, Pediatric, Adult Ventilator

Condition: Pre-owned, Calibrated, Patient Ready

TECHNICAL DATA EVITA XL
Patient type – Adults, children, infants (body weight of at least 3 kg/6.6 lbs)
– Premature infants with NeoFlow option
Ventilation settings
Ventilation mode – IPPV, IPPVAssist/CMV, CMVAssist
– SIMV, SIMVPsupp
– MMV, MMVPsupp
– BIPAP1)
, BIPAP1)
ASB, BIPAP1)
Assist / PCV+, PCV+Psupp, PCV+Assist
– APRV
– CPAP, CPAPASB / CPAP/Psupp, CPAP/Psupp
– ILV
– PPS (optional)
Enhancements – AutoFlow™ – Automatic adaptation of inspiratory flow
in volume controlled modes
– ATC™ – Automatic Tube Compensation™
– NIV – Mask Ventilation (optional)
– SmartCare®/PS– Automated clinical protocol
in CPAP/ASB / CPAP/Psupp (optional)
– Lung Protection Package –
Recruitment maneuvre and Low Flow maneuvre (optional)
Ventilation frequency (f) 0 to 100/min, 0 to 150/min (Neonatal)
Inspiration time (Tinsp) 0.1 to 10 s
Tidal volume (VT) (BTPS*) – 0.1 to 2.0 L (Adult) / 0.02 to 0.3 L (Pediatric)
– 0.003 to 0.1 L (Neonatal)
Inspiratory flow – 6 to 120 L/min (Adult)
– 6 to 30 L/min (Pediatric and Neonatal)
Inspiratory pressure 0 to 95 mbar/cmH2O
PEEP / intermittent PEEP 0 to 50 mbar/cmH2O
Pressure assist ASB/Psupp 0 to 95 mbar/cmH2O
Rise time for inspiratory pressure 0 to 2 s
O2 concentration 21 to 100 Vol.%
Multi-sense Trigger Criteria Internal automatic pressure trigger, Flow, Volume
(Flow adjustable 0.3 to 15 L/min)
Measured values displayed
Airway pressure Peak pressure, plateau pressure, mean pressure,
PEEP, min. pressure (-45 to 110 mbar/cmH2O)
Minute volume (MV), (BTPS*) MV, MVspont (0 to 120 L/min, MVleak (0 to 99 L/min)
Tidal volume (VT), (BTPS*) Vtasb 0 - 10, respectively 0 - 3999 ml02 | DRÄGER EVITA XL
Breathing frequency (f) ftotal, fspon, fmand. (0 to 300 bpm)
O2 concentration (FiO2) Inspired O2 concentration (15 to 100 Vol.%)
Lung mechanics – Resistance (0 to 600 mbar/cmH2O L/s)
– Compliance (0 to 300 mL/mbar/cmH2O)
Breathing gas temperature 18 °C to 51 °C
Capnography (etCO2) (optional) – 0 to 100 mmHg
CO2 production (VCO2) – 0 to 999 mL/min, STPD*
Serial dead space Vds – 0 to 999 mL, BTPS*
Dead space ventilation (Vds/VT) – 0 to 99%
Weaning parameters – RSB (0 to 9999 (min x L)) / NIF (-45 to 0 mbar/cmH2O)
Alarms / Monitoring
Airway pressure High / Low
Expired minute volume High / Low
Tidal volume High
Apnea alarm Time 5 to 60 s
Spontaneous breath frequency High
Inspired O2 concentration High / Low
Breathing gas temperature High
SpO2 pulse (optional) High / Low
etCO2 (optional) High / Low
Performance data
Valve response time T0...90 ≤ 5 ms
Control principle Time cycled, volume constant, pressure-controlled
Safety relief valve 100 mbar/cmH2O
Leakage and hose system compensation compliance automatic
Max. flow for pressure support and spontaneous breathing 180 L/min
Outlet for pneumatic nebulizer
Operating data
Mains power connection 100 to 240 V, 50/60 Hz, 10 to 30 V DC
Power consumption Approx. 125 W
Gas supply operating pressure O2, air: 2.7 to 6 bar / 39 to 87 PSI
Physical specifications
Dimensions ventilator (W x H x D) 530 x 315 x 450 mm / 20.9 x 12.4 x 17.7 inches (without trolley)
Diagonal screen size 15" TFT color touch screen
Weight basic unit Approx. 50 kg / 64 lbs
Machine outputs:
Digital output Output and reception via an RS 232 C interface
Digital output Output for independent lung ventilation (ILV)
Digital output (optional) For output and reception via two RS 232 C interfaces
Analog output (optional) For analog output of two measured values